Short echo time MR spectroscopic imaging for neonatal pediatric imaging

Dong Hyun Kim, A. J. Barkovich, D. B. Vigneron

Research output: Contribution to journalArticlepeer-review

8 Citations (Scopus)

Abstract

A short echo time (30 milliseconds) MR spectroscopic imaging pulse sequence was implemented for applications of neonatal brain imaging. Multiple spatial saturation bands were used to eliminate strong signals originating from the subcutaneous lipids to enable volumetric region coverage. Metabolite signal intensity-to-noise ratio ≤40 was acquired in 9 minutes of scan time over an 8 x 8 x 8 spatial matrix with 1 cm3 isoresolution.

Original languageEnglish
Pages (from-to)1370-1372
Number of pages3
JournalAmerican Journal of Neuroradiology
Volume27
Issue number6
Publication statusPublished - 2006 Jun