Scleral fixation of a silicone plate haptic intraocular lens in a patient showing repeated pupillary intraocular lens capture

Kyoung Tak Ma, Sung Soo Kim, Gong Jae Seong, Hyung Keun Lee, Chan Yun Kim

Research output: Contribution to journalArticlepeer-review

4 Citations (Scopus)

Abstract

A 39-year-old man with Marfan syndrome underwent scleral fixation using a three-piece foldable intraocular lens and experienced repeated pupillary intraocular lens captures secondary to a pliable iris. As a result, the patient needed an additional pupilloplasty. In the opposite eye, a silicone plate haptic intraocular lens was used for scleral fixation because it was thought that the plate design could prevent pupillary intraocular lens capture, even in cases involving a pliable iris. At the 27-month follow-up examination, the patient showed good centration of the intraocular lens in the eye that used the silicone plate haptic intraocular lens without pupillary intraocular lens capture.