Quantized hall resistance measurement at ksri

Kyung Hwa Yoo, Kwang Pill Lee, Kwang Min Yu, Jong Chul Park

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

This paper describes the measurement system of quantized Hall resistances used at the Korea Standards Research Institute (KSRI) and reports the results obtained with GaAs-AIGaAs heterostructures. The result in terms of Ω KSRI ’ which has been maintained as a legal unit until 1989, is RH. 4= 6453.201 75 (l + 0.38 X 10 –6 ± 0.39 x 10 –6) Ω KSRI on 1990-1-1 and we obtain Ω KSRI = (l-0.38 × 10– 6 ± 0.39 × 10–6) Ω.

Original languageEnglish
Pages (from-to)253-255
Number of pages3
JournalIEEE Transactions on Instrumentation and Measurement
Volume40
Issue number2
DOIs
Publication statusPublished - 1991 Apr

All Science Journal Classification (ASJC) codes

  • Instrumentation
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Quantized hall resistance measurement at ksri'. Together they form a unique fingerprint.

Cite this