Quantitative evaluation of acquisition parameters in three-dimensional imaging with multidetector computed tomography using human skull phantom.

Dong Ook Kim, Hee Joung Kim, Haijo Jung, Ha Kyu Jeong, Soon Il Hong, Kee Deog Kim

Research output: Contribution to journalArticlepeer-review

22 Citations (Scopus)

Abstract

This study evaluated the quantitative measurements of three-dimensional (3D) volume images using multidetector row computed tomography (MDCT) and one skull specimen. Twenty-one linear distances were measured five times each by vernier caliper. A dry human skull was imaged with MDCT for various acquisition parameters at slice thicknesses of 1.25, 2.50, 3.75, and 5.00 mm for the different acquisition modes of axial and spiral scan. The distance of each corresponding item displayed on 3D rendered images was measured seven times by an uninvolved observer using a 3D software tool. Data analysis was performed to determine if there were any statistically significant differences in acquisition parameters. No significant image differences were found among the scan modes for each slice thickness. For a given scan mode, acquisition slice thickness was the important factor for quantitative measurement of 3D rendered CT images.

Original languageEnglish
Pages (from-to)254-257
Number of pages4
JournalJournal of Digital Imaging
Volume15 Suppl 1
DOIs
Publication statusPublished - 2002

All Science Journal Classification (ASJC) codes

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Computer Science Applications

Fingerprint