Abstract
A new rigid graphite-epoxy composite electrode for electrochemical detection in microchip electrophoresis is described. The end-column wall-jet detection design relying on tubular rigid composite working electrode shows favorably low noise level and high signal-to-noise ratio when compared to glassy carbon detector. The performance of rigid graphite-epoxy composite detector was compared to glassy carbon detector using dopamine and catechol as model analytes. The various parameters of the microchip electrophoresis-rigid graphite-epoxy composite detector were optimized.
Original language | English |
---|---|
Pages (from-to) | 207-210 |
Number of pages | 4 |
Journal | Electroanalysis |
Volume | 18 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2006 Jan |
All Science Journal Classification (ASJC) codes
- Analytical Chemistry
- Electrochemistry