We report a breath hydrogen analyzer based on Pd-coated SnO2 nanorods (Pd-SnO2 NRs) sensor integrated into a miniaturized gas chromatography (GC) column. The device can measure a wide range of hydrogen (1–100 ppm), within 100 s, using a small volume of human breath (1 mL) without pre-concentration. Especially, the mini-GC integrated with Pd-SnO2 NRs can detect 1 ppm of H2, as a lower detection limit, at a low operating temperature of 152◦C. Furthermore, when the breath hydrogen analyzer was exposed to a mixture of interfering gases, such as carbon dioxide, nitrogen, methane, and acetone, it was found to be capable of selectively detecting only H2. We found that the Pd-SnO2 NRs were superior to other semiconducting metal oxides that lack selectivity in H2 detection. Our study reveals that the Pd-SnO2 NRs integrated into the mini-GC device can be utilized in breath hydrogen analyzers to rapidly and accurately detect hydrogen due to its high selectivity and sensitivity.
Original language | English |
---|---|
Article number | 2056 |
Journal | Sensors |
Volume | 22 |
Issue number | 5 |
DOIs | |
Publication status | Published - 2022 Mar 1 |
Bibliographical note
Publisher Copyright:© 2022 by the authors. Licensee MDPI, Basel, Switzerland.
All Science Journal Classification (ASJC) codes
- Analytical Chemistry
- Information Systems
- Atomic and Molecular Physics, and Optics
- Biochemistry
- Instrumentation
- Electrical and Electronic Engineering