Evaluation of the extraction method for the cytotoxicity testing of latex gloves

Hyun Sook Baek, Ja Young Yoo, Dong Kyun Rah, Dong Wook Han, Dong Hee Lee, Oh Hun Kwon, Jong Chul Park

Research output: Contribution to journalArticlepeer-review

48 Citations (Scopus)

Abstract

In this study, the cytotoxicity of medical latex gloves to cultured L-929 cells was determined using various extraction conditions. According to the extraction time and temperature, three types of extraction conditions were used: 1) 24 h at 37 °C; 2) 72 h at 37 °C; 3) 72 h at 50 °C. Also, four different extraction vehicles were used, namely, distilled water (DW), 9 g/l sodium chloride (saline) in DW, and culture media with or without serum. Under the above-mentioned conditions, the samples were extracted and then 2-fold serially diluted in the concentration range 3.13 - 50%. When extracted with either DW or saline for 24 h or 72 h at 37 °C, only 50% diluted samples showed distinct cytotoxicity to L-929 cells. Moreover, no cytotoxic potentials were observed when gloves were extracted with DW or saline at 50 °C for 72 h. Cytotoxicity was markedly greater when gloves were extracted with culture medium, irrespective of the presence of serum in the medium. These results suggest that optimal extraction conditions should be established for the cytotoxicity evaluations of biomaterials and medical devices.

Original languageEnglish
Pages (from-to)579-583
Number of pages5
JournalYonsei medical journal
Volume46
Issue number4
DOIs
Publication statusPublished - 2005 Aug 31

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Fingerprint

Dive into the research topics of 'Evaluation of the extraction method for the cytotoxicity testing of latex gloves'. Together they form a unique fingerprint.

Cite this