TY - JOUR
T1 - Effect and Safety of Pressure Sensor-equipped Handpiece in Phacoemulsification System
AU - Kim, Jin Yeong
AU - Kim, Harin
AU - Jun, Ikhyun
AU - Kim, Tae Im
AU - Seo, Kyoung Yul
N1 - Publisher Copyright:
© 2023 The Korean Ophthalmological Society.
PY - 2023/10
Y1 - 2023/10
N2 - Purpose: To evaluate the effect of the Active Sentry handpiece of the Centurion Vision System compared to the Centurion Ozil handpiece for phacoemulsification in cataract surgery. Methods: A retrospective study was conducted on 281 patients (449 eyes) who underwent cataract surgery between August 2020 and June 2021. Preoperative measurements, intraoperative parameters, complication rate, and postoperative outcomes were compared between the Active Sentry handpiece and the Centurion Ozil handpiece groups. Additionally, the parameters were compared in different cataract severity groups and multiple predictive factors for the number of active surge mitigation (ASM) actuations were assessed with the Active Sentry handpiece. Results: There were 198 eyes in the Active Sentry group and 251 eyes in the Centurion Ozil group. There were no statistically significant differences between the two groups, as the cumulative dissipated energy in the Active Sentry and Centurion Ozil groups were 8.32 ± 7.74 and 7.87 ± 9.25 μJ, respectively (p = 0.576). Total surgery time, ultrasound usage time, aspiration time, amount of fluid aspirated, postoperative corrected distant visual acuity, and postoperative decrease in corneal endothelial cell density were comparable between the two groups. The significant contributors to the number of ASM actuations were age, preoperative corrected distant visual acuity, axial length, and total ultrasound time. Conclusions: There was no clear advantage of the Active Sentry handpiece compared to the Centurion Ozil handpiece. ASM actuation increases with age, poor visual acuity before surgery, short axial length, and prolonged ultrasound usage time. It is expected that in more severe and high-risk cataract surgery, the Active Sentry handpiece functions more effectively, possibly affecting the safety and prognosis.
AB - Purpose: To evaluate the effect of the Active Sentry handpiece of the Centurion Vision System compared to the Centurion Ozil handpiece for phacoemulsification in cataract surgery. Methods: A retrospective study was conducted on 281 patients (449 eyes) who underwent cataract surgery between August 2020 and June 2021. Preoperative measurements, intraoperative parameters, complication rate, and postoperative outcomes were compared between the Active Sentry handpiece and the Centurion Ozil handpiece groups. Additionally, the parameters were compared in different cataract severity groups and multiple predictive factors for the number of active surge mitigation (ASM) actuations were assessed with the Active Sentry handpiece. Results: There were 198 eyes in the Active Sentry group and 251 eyes in the Centurion Ozil group. There were no statistically significant differences between the two groups, as the cumulative dissipated energy in the Active Sentry and Centurion Ozil groups were 8.32 ± 7.74 and 7.87 ± 9.25 μJ, respectively (p = 0.576). Total surgery time, ultrasound usage time, aspiration time, amount of fluid aspirated, postoperative corrected distant visual acuity, and postoperative decrease in corneal endothelial cell density were comparable between the two groups. The significant contributors to the number of ASM actuations were age, preoperative corrected distant visual acuity, axial length, and total ultrasound time. Conclusions: There was no clear advantage of the Active Sentry handpiece compared to the Centurion Ozil handpiece. ASM actuation increases with age, poor visual acuity before surgery, short axial length, and prolonged ultrasound usage time. It is expected that in more severe and high-risk cataract surgery, the Active Sentry handpiece functions more effectively, possibly affecting the safety and prognosis.
KW - Active fluidics
KW - Active sentry handpiece
KW - Cataract
KW - Phacoemulsification
KW - Retrospective study
UR - http://www.scopus.com/inward/record.url?scp=85178914282&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85178914282&partnerID=8YFLogxK
U2 - 10.3341/kjo.2022.0157
DO - 10.3341/kjo.2022.0157
M3 - Article
AN - SCOPUS:85178914282
SN - 1011-8942
VL - 37
SP - 387
EP - 394
JO - Korean journal of ophthalmology : KJO
JF - Korean journal of ophthalmology : KJO
IS - 5
ER -