Development of a PMMA phantom as a practical alternative for quality control of gamma knife® dosimetry

Jae Pil Chung, Young Min Seong, Tae Yeon Kim, Yona Choi, Tae Hoon Kim, Hyun Joon Choi, Chul Hee Min, Hamza Benmakhlouf, Kook Jin Chun, Hyun Tai Chung

Research output: Contribution to journalArticlepeer-review

11 Citations (Scopus)

Abstract

Background: To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom. Methods: A multi-purpose PMMA phantom was developed to measure the absorbed dose rate to water and the dose distribution of a GK. The phantom consists of a hemispherical outer phantom, one exchangeable cylindrical chamber-hosting inner phantom, and two film-hosting inner phantoms. The radius of the phantom was determined considering the electron density of the PMMA such that it corresponds to 8 g/cm 2 water depth, which is the reference depth of the absorbed dose measurement of GK. The absorbed dose rate to water was measured with a PTW TN31010 chamber, and the dose distributions were measured with radiochromic films at the calibration center of a patient positioning system of a GK Perfexion. A spherical water-filled phantom with the same water equivalent depth was constructed as a reference phantom. The dose rate to water and dose distributions at the center of a circular field delimited by a 16-mm collimator were measured with the PMMA phantom at six GK Perfexion sites. Results: The radius of the PMMA phantom was determined to be 6.93 cm, corresponding to equivalent water depth of 8 g/cm 2 . The absorbed dose rate to water was measured with the PMMA phantom, the spherical water-filled phantom and a commercial solid water phantom. The measured dose rate with the PMMA phantom was 1.2% and 1.8% higher than those measured with the spherical water-filled phantom and the solid water phantom, respectively. These differences can be explained by the scattered photon contribution of PMMA off incoming 60 Co gamma-rays to the dose rate. The average full width half maximum and penumbra values measured with the PMMA phantom showed reasonable agreement with two calculated values, one at the center of the PMMA phantom (LGP6.93) and other at the center of a water sphere with a radius of 8 cm (LGP8.0) given by Leksell Gamma Plan using the TMR10 algorithm. Conclusions: A PMMA phantom constructed in this study to measure the absorbed dose rates to water and dose distributions of a GK represents an acceptable and practical alternative for GK dosimetry considering its cost-effectiveness and ease of handling.

Original languageEnglish
Article number176
JournalRadiation Oncology
Volume13
Issue number1
DOIs
Publication statusPublished - 2018 Sept 14

Bibliographical note

Publisher Copyright:
© 2018 The Author(s).

All Science Journal Classification (ASJC) codes

  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Development of a PMMA phantom as a practical alternative for quality control of gamma knife® dosimetry'. Together they form a unique fingerprint.

Cite this