Abstract
Background: FLAIR (fluid attenuated inversion recovery) imaging via synthetic MRI methods leads to artifacts in the brain, which can cause diagnostic limitations. The main sources of the artifacts are attributed to the partial volume effect and flow, which are difficult to correct by analytical modeling. In this study, a deep learning (DL)-based synthetic FLAIR method was developed, which does not require analytical modeling of the signal. Purpose: To correct artifacts in synthetic FLAIR using a DL method. Study Type: Retrospective. Subjects: A total of 80 subjects with clinical indications (60.6 ± 16.7 years, 38 males, 42 females) were divided into three groups: a training set (56 subjects, 62.1 ± 14.8 years, 25 males, 31 females), a validation set (1 subject, 62 years, male), and the testing set (23 subjects, 57.3 ± 20.4 years, 13 males, 10 females). Field Strength/Sequence: 3 T MRI using a multiple-dynamic multiple-echo acquisition (MDME) sequence for synthetic MRI and a conventional FLAIR sequence. Assessment: Normalized root mean square (NRMSE) and structural similarity (SSIM) were computed for uncorrected synthetic FLAIR and DL-corrected FLAIR. In addition, three neuroradiologists scored the three FLAIR datasets blindly, evaluating image quality and artifacts for sulci/periventricular and intraventricular/cistern space regions. Statistical Tests: Pairwise Student's t-tests and a Wilcoxon test were performed. Results: For quantitative assessment, NRMSE improved from 4.2% to 2.9% (P < 0.0001) and SSIM improved from 0.85 to 0.93 (P < 0.0001). Additionally, NRMSE values significantly improved from 1.58% to 1.26% (P < 0.001), 3.1% to 1.5% (P < 0.0001), and 2.7% to 1.4% (P < 0.0001) in white matter, gray matter, and cerebral spinal fluid (CSF) regions, respectively, when using DL-corrected FLAIR. For qualitative assessment, DL correction achieved improved overall quality, fewer artifacts in sulci and periventricular regions, and in intraventricular and cistern space regions. Data Conclusion: The DL approach provides a promising method to correct artifacts in synthetic FLAIR. Level of Evidence: 4. Technical Efficacy: Stage 1. J. Magn. Reson. Imaging 2019;50:1413–1423.
Original language | English |
---|---|
Pages (from-to) | 1413-1423 |
Number of pages | 11 |
Journal | Journal of Magnetic Resonance Imaging |
Volume | 50 |
Issue number | 5 |
DOIs | |
Publication status | Published - 2019 Nov 1 |
Bibliographical note
Publisher Copyright:© 2019 International Society for Magnetic Resonance in Medicine
All Science Journal Classification (ASJC) codes
- Radiology Nuclear Medicine and imaging