Abstract
Medical linear accelerators operating above 10 MV require door shielding for neutrons in addition to photons. A criterion for choice of optimal configuration of lamination of BPE (Borated Polyethylene) and lead is not clear. Moreover, optimal configuration cannot be determined by the conventional method using an analytical formula and simple measurement. This study performs Monte Carlo simulation of radiation field in a commercial LINAC room with 15 MV X-ray sources. Considering two configuration of lamination such as 'lead-BPE' and 'lead-BPE-lead', dose equivalents are calculated by using the MCNPX code and comparative analyses are performed with each other. The obtained results show that there is no significant difference in neuron shielding between both configurations, whereas lead-BPE-lead is more effective for photon shielding. It is also noted that the absolute values of neutron doses are much greater than that of photon doses outside as well as inside the door, by three orders of magnitude. As a conclusion, the laminating of lead-BPE is suggested as the optimal configuration from the viewpoint of simplicity in fabrication and handling, even though it has no significant difference from lead-BPE-lead in terms of total dose equivalent.
Original language | English |
---|---|
Pages (from-to) | 50-53 |
Number of pages | 4 |
Journal | Journal of Nuclear Science and Technology |
Volume | 45 |
DOIs | |
Publication status | Published - 2008 |
Bibliographical note
Publisher Copyright:© 2008, Taylor and Francis Ltd. All rights reserved.